All LCI scripts share six parameters.
| Name | type | valid values | comment |
| number | parameter | int | Number of events in innermost loop |
| period | parameter | int | Time between events (ticks) |
| delay | parameter | int | Time between DAC values |
| latc_delay | parameter | int | Time after LATC configuration |
| strobe | parameter | ON,OFF | ON means inject charge |
| zero_suppress | parameter | ON,OFF | ON means apply zero-suppression |
These are all specified in a single XML node <lci_parameters> which is placed as the inner-most node of the loop definition
<lci_loop loop_id='name'>
<lci_parameters number='50' period='200000' latc_delay='0' delay='2'
strobe='ON' zero_suppress='OFF'/>
</lci_loop>